论文部分内容阅读
对六台非零位输出的TRG-1型潮汐重力仪进行摆位与标定常数的关系的实验。通过仪器对标定脉冲的响应的研究,确定各台仪器灵敏度和格值随摆位变化的关系函数。发现用变格值处理数据可以有效地消除摆位对潮汐因子测定的影响。
Experiment on the relationship between setting and calibration constant of six non - zero TRG - 1 tidal gravimeters. Through the instrument calibration of the pulse response to determine the sensitivity of each instrument and the value of the grid with the changing position of the relationship function. It is found that processing the data with the modified values can effectively eliminate the influence of setting on the tide factor measurement.