海珠区区属医用X线机防护情况的调查分析

来源 :广州医药 | 被引量 : 0次 | 上传用户:cooltom
下载到本地 , 更方便阅读
声明 : 本文档内容版权归属内容提供方 , 如果您对本文有版权争议 , 可与客服联系进行内容授权或下架
论文部分内容阅读
1987年,我们根据上级的指示,对我区区属医疗单位的医用X线机的防护情况进行全面调查,并对防护区进行了测试,为进一步搞好X线机防护提供了依据。现将结果报告如下。一、一般情况 1.设备情况:我区区属医疗单位18间,医用X线机共26台(国产23台,进口1台、杂牌老机2台)。其中30mA以下13台,50mA4台,100mA1台,200mA以上8台。从事放射工作28人。 2.机房的使用面积:小于100mA的机房平均面积为12.21m~2;大于200mA的机房平均面积为23.57m~2。 In 1987, according to instructions from the superiors, we carried out a comprehensive investigation on the protection of medical X-ray machines of medical units in our district and conducted tests on the protected areas to provide a basis for further improving the protection of X-ray machines. The results reported below. First, the general situation 1. Equipment: My area is a medical unit 18, a total of 26 sets of medical X-ray machine (domestic 23, imported 1 sets, 2 sets of old machines). Including 30mA following 13, 50mA4 Taiwan, 100mA1 Taiwan, 200mA above 8. 28 people engaged in radiation work. 2. The use of room area: less than 100mA average room area of ​​12.21m ~ 2; greater than 200mA average room area of ​​23.57m ~ 2.
其他文献
文章给出电阻盒式静电偏转器场分布和象差的简单计算公式。这种静偏器的场分布与选取合适“调节参数”的多极静偏器的场分布具有相同分布模式。经简化后,十六个象差系数公式
知识经济时代呼唤知识创新型的高新技术企业,呼唤知识创新型的企业家。知识创新型的企业经营者,特别是高新技术企业经营者需要创新的经营体制来实现其经营管理创新,从而促进
QJ33型万能比例臂可以用直流比较仪电位差计和三次平衡电桥等测量电阻的方法进行检定,但这种方法操作和计算均麻烦,给检定工作带来困难。本文介绍用一般省级计量部门检定高
供热、热水采暖及空气调节水系统中,热(冷)媒的水力输配系统往往存在水力失调。这意味着热(冷)量输配不均匀,导致部分房间过热或过冷,同时浪费能量。本文介绍平衡阀及其智能
请下载后查看,本文暂不支持在线获取查看简介。 Please download to view, this article does not support online access to view profile.
饲料生产企业因其自身特点在财务处理上有许多不规范之处,应按会计谨慎性原则,对任何账务处理都从实质入手,以提供正确的会计信息。 Because of its own characteristics, f
随着信息化社会的发展,培养创新型人才,变得尤为重要。在中学物理教学中实施创新教育,培养学生的科学创新精神和提高创新能力,是落实素质教育的核心。但就目前教学的实际情况
一、注意膳食平衡多摄取胶质和蛋白质,为组织细胞的新生提供必要的营养物质,补充肌肤维持活力所必需的能量。如肉类、鱼类、蛋类。切忌盲目节食减肥,以致营养不良,影响皮肤弹
我队曾使用LN-1型、唐山式、液压马达式等类型拧管机。为拧管机打叉的防甩问题,曾使用过弹簧安全钩、弹簧卡块、弹簧卡球等安全装置,均收到一定效果。但也存在着防甩失灵、操作
这里所选的是三封名人的家书,三封信中都流露出写信者内心深深的爱、浓浓的情。更让人感动的是写信者在信中表现出对收信人成长的关注:奥巴马希望女儿们要学会担负责任,林肯